☢️ Theremino Radon Ion Chamber V7
The Theremino Radon Ion Chamber V7 represents the culmination of nearly a decade of research, introducing substantial improvements in stability, precision, and ease of construction for high-sensitivity environmental radioactivity measurement, particularly Radon gas detection.
Key Performance Enhancements
-
Superior Parameter Stability: Achieves extremely stable measurement parameters, even under strong temperature variations, solving a major instability issue found in earlier designs.
-
Greatly Improved Precision: The overall measurement accuracy and resolution have been significantly enhanced.
-
Reduced Noise Floor: Features a lower intrinsic background noise and reduced sensitivity to external electrical disturbances.
-
Optimized Electronics Layout:
-
The entire electronics package (PCB modules) is now unified on the same side of the chamber, eliminating the need for long, shielded cables that previously ran to an amplifier on the opposite end.
-
Easy Accessibility: The design allows for measurements to be taken on components even when the chamber is open (by covering the FET with a rectangular shield and the hole with a fine brass mesh), simplifying testing and calibration.
-
Simplified Construction: The new compact PCB modules can be secured using rigid tinned copper wires, simplifying assembly and module replacement.
-
Reliability and Electrical Stability Solutions
Version 7 integrates critical hardware fixes to ensure reliable long-term performance:
-
Humidity Resilience: While extreme humidity remains a challenge for any ion chamber, the V7’s design minimizes sensitivity loss caused by moist air.
-
High-Voltage Stability Fixes: Includes hardware corrections (like increasing external capacitors on the voltage regulator and adding a capacitor on the input) to eliminate self-oscillation and low-frequency “wobble” on the High Voltage power supply. This ensures the signal baseline is stable and eliminates the risk of spurious pulse counts.
-
Central Electrode Design: Specifies the use of a central electrode made from a brass tube (instead of a stretched wire) to significantly improve resistance to mechanical noise and enhance efficiency.